Introducing Nanotrac Wave
Microtrac has been a pioneer in particle sizing technology for over 30 years! The Nanotrac Wave is the latest generation of sub-micron Particle Size and Zeta Potential analyzers from Microtrac! The enhanced design on the Nanotrac Wave features faster measurements, smaller particle size capability (to 0.8 nm), higher precision, higher accuracy and advanced software capabilities in a small, robust dynamic light scattering instrument with no moving components. Nanotrac Wave complies with the ISO 13321 standard for Dynamic Light Scattering. |

Nanotrac wave |
|
Features |
 |
|
Technology |
Dynamic Light Scattering incorporating the patented Controlled Reference Method for
advanced power spectrum analysis of Doppler shifts under Brownian Motion. |
Range |
Measurement capability from 0.8 to 6500 nanometers. Zeta Potential from -125 to +125MV |
Sample Size |
Typically less than 3ml in standard cell, or 0.2ml with small cell option. |
Rapid Analysis |
15 to 30 second analysis times in most cases. |
Analysis Simplicity |
No “A Priori” or advance knowledge of the sample is required. Nanotrac will accurately
report monomodal, multimodal, broad and narrow distributions without the need to select
special calculation algorithms. |
Accuracy |
Measures using Mie scattering calculations for spherical particles and proprietary Mie
calculations for non-spherical particles. This feature is unique to Microtrac. |
Repeatability |
Better than 1% from instrument to instrument for 100nm polystyrene. |
Traceability |
Particle Size measurements traceable to NIST standards. |
Sample Cell Modules |
Nano, Zeta, Micro insert for Nano Cell, Flow thru cell with or withouit cap. |
Concentration |
Capable to measure up to 40% by solids in some cases. This feature reduces the need to
dilute samples which may affect particle size or colloid stability. Low concentration
measurements can be as low as 0.1ppm for 200nm polystyrene. |
Temperature Control |
Precise temperature control using a Peltier control device. |
Versatility |
Nanotrac Wave probe can be external for use in Dip-N-Run or in-line applications. Nanotrac
Wave is also compatible with the Zetrator titration device. |
Security/Validation |
Nanotrac Wave FLEX software is compatible with FDA 21 CFR Part 11 protocols. Full
IQ/OQ/PQ validation documentation is available for Nanotrac Wave. |
|
Specifications |
 |
|
Particle Size |
0.08 to 6500 nanometers (displayed in nanometers, microns or angstroms) |
Zeta Potential |
-125 to +125 mV, Mobility -10 to +10 microseconds/volt/cm, size range 10nm to 20 microns |
Molecular Weight |
1kDa to 20 MDa, Hydrodynamic, Debye Plot with 2nd Virial co-efficient calculation |
Sample Volume |
Size only - 50 ul to 500 ul, with micro insert min 10 to 20 ul. Zeta 0.7 to 3 ml. |
Measurement Angle |
180 Degrees |
Repeatability |
1% or better for 100nm Polystyrene |
Concentraion Limits |
From ppb to 40% by volume in certain conditions. |
Laser |
Laser Diode, 780 nm , 3mW Nominal, Class IIIB, no alignment required. |
Temperature Control |
5 to 90 Degrees Centigrade +/- 0.2 Degrees using Peltier Device |
Sample Cell Modules |
Nano, Zeta, Micro insert for Nano Cell, Flow thru cell with or withouit cap. |
Chemical Compatibility |
Aqueous, polar and non polar solvents, organic solvents. Surfaces-Stainless steel, sapphire and Teflon |
pH Range |
2 to 12 pH |
Conductivity Range |
1 to 200 mS/cm |
Environmental |
Operating Temperature 10 to 50 Degrees C
Humidity up to 90% non condensing |
Dimensions |
15L x 14W x 13H inches (38.1 x 35.5 x 33 cms) |
Weight |
30 Lbs, approx 15 Kg |
Electrical |
90 to 240 VAC, 47 to 63 HZ, 5 Amps |
|
|